PN 9000-133-VE, Rev U
Replacing the Sensor Board
To replace the
Sensor
board (Figure 3-10):
1. Disconnect all wires.
2. Loosen the two Phillips screw located on the back of the electronics plate.
3. Slide the board out until the pegs at the top line up with the holes in the
Sensor
board plate and drop it down.
4. Replace the
Sensor
board.
5. Properly tighten the screws.
3-28 | Thermox
®
WDG-V / VC / VCM / VM
Summary of Contents for WDG-V
Page 12: ...PN 9000 133 VE Rev U This page intentionally left blank xii Thermox WDG V VC VCM VM...
Page 60: ...PN 9000 133 VE Rev U Figure 2 16 RCU Connections 2 34 Thermox WDG V VC VCM VM...
Page 93: ...PN 9000 133 VE Rev U Figure 3 4 Thermocouple replacement Maintenance and Troubleshooting 3 19...
Page 97: ...PN 9000 133 VE Rev U Figure 3 6 Flow sensor replacement Maintenance and Troubleshooting 3 23...
Page 103: ...PN 9000 133 VE Rev U Figure 3 10 Sensor board removal Maintenance and Troubleshooting 3 29...