Installing the Analyzer
• The analyzer ambient temperature range is -25 °C to 65 °C.
• For purged units, the analyzer ambient temperature range is -20 °C to
60 °C.
• When installing a probe heater or exhaust tube along with the analyzer,
attach it to the back of the analyzer before mounting the analyzer to the
process.
Do not open a Z-purge, hinged-analyzer enclosure door until you first
verify that the area has been classified as nonhazardous.
Methods for Mounting the Analyzer
There are three methods for mounting a WDG-V Analyzer to the process. Each
method is explained in the following pages. Mounting dimensions for the
hinged style enclosure, including the Z-purge option, are provided in Figure
2-5.
If the protection nipple provided by AMETEK extends past the inside
of the refractory wall on a process that has a gas temperature over
650 °C (1200 °F), the nipple should be cut down before installation to
ensure it remains flush with the inside wall.
Installation and Start-Up
|
2-11