DC Power Supplies
ReFlex Power™
6-62
M380056-01 Rev L
11. Turn off the output with the following command:
*RST[n]
S
ERIES
C
ONFIGURATION
O
PERATION
C
HECK
1. Before attempting to verify series operation, perform the power-on
test of the individual modules per the section, Power-On Check,
and then the standalone operation test per section, Standalone
Configuration.
2. Connect the outputs for series operation per the section, Series
Configuration. Ensure that the outputs and remote sense are
appropriately hardwired prior to issuing any of the series-
connection SCPI commands.
3. Connect appropriate instrumentation for measuring the output
voltage and current, as well as a load.
4. Configure the modules into a series group with the following
command, where [n1] is the master module of the group. Although
the example is for two modules, it could be extended to a greater
number by listing additional parameter slot numbers,
[n]
.
SYST:GRO:DEF:SER [n1],[n2]
5. Verify that the series group has been properly set up with the
following command. The response to the query is a listing of the
series group, the master module of the series group, and the slave
module of the series group. For example, a group of modules with
a master in slot 3 and a slave in slot 4 would return the following
parameters: 2003,1003,1004. The parameter,
[2000+n1]
, is the
sum of 2000 and the slot number of the master module.
SYST:GRO:CAT:SER?
6. Turn on the isolation and remote sense relays, turn on the output
converters, and program full-scale voltage and current with the
following commands. The parameter,
<fval>
, is the full-scale
output voltage or current of the particular model under test. The
parameter,
[2000+n1]
, is the sum of 2000 and the slot number of
the master module. For example, if the master module is located
in slot 1, the parameter value would be 2001.
OUTP[2000+n1]:ISOL 1
OUTP[2000+n1]:SENS 1
OUTP[2000+n1]:STAT 1
SOUR[2000+n1]:CURR <fval>
SOUR[2000+n1]:VOLT <fval>
7. Measure the output voltage and current with the external
instruments.
Summary of Contents for Elgar ReFlex Power
Page 1: ...M380056 01 Rev L www programmablepower com ReFlex Power Operation Manual...
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank...
Page 8: ...iv This page intentionally left blank...
Page 10: ...vi This page intentionally left blank...
Page 21: ...M380056 01 Rev L xvii This page intentionally left blank...
Page 22: ......
Page 85: ...ReFlex Power Controller ReFlex Power 3 22 M380056 01 Rev L This page intentionally left blank...
Page 89: ...AC Power Supplies ReFlex Power 4 4 M380056 01 Rev L Figure 4 1 ACPS Module Front Panel 875VA...
Page 120: ...ReFlex Power AC Power Supplies M380056 01 Rev L 4 35 Figure 4 11 Parallel Output Configuration...
Page 147: ...AC Power Supplies ReFlex Power 4 62 M380056 01 Rev L RST n...
Page 157: ...AC Power Supplies ReFlex Power 4 72 M380056 01 Rev L 26 Reset the unit RST n...
Page 274: ...ReFlex Power DC Power Supplies M380056 01 Rev L 6 71 18 Turn off the output RST n...
Page 275: ...DC Power Supplies ReFlex Power 6 72 M380056 01 Rev L This page intentionally left blank...
Page 279: ...Active Loads ReFlex Power 7 4 M380056 01 Rev L Figure 7 1 LPAL 375W Front Panel...
Page 280: ...ReFlex Power Active Loads M380056 01 Rev L 7 5 Figure 7 2 HPAL 750W Front Panel...
Page 281: ...Active Loads ReFlex Power 7 6 M380056 01 Rev L Figure 7 3 Typical Active Load Rear Panel...
Page 359: ...Active Loads ReFlex Power 7 84 M380056 01 Rev L This page intentionally left blank...