Recommended Separation Distances Between Portable and Mobile RF Communication

Equipment and system.

The system is intended for use in an electromagnetic environment in which radiated RF 

disturbances are controlled. The user of the system can help prevent electromagnetic 

interference by maintaining a minimum distance between portable and mobile RF 

communications equipment (transmitters and the system as recommended below, 

according to the maximum output power of the communication equipment.

Rated maximum 
output power (W) 
of transmitter

Separation distance (m) according to frequency of transmitter

150kHZ to 80MHz
D = 1.17√P

80MHz to 800MHZ
D = 1.17√P

800MHz to 2.7GHz
D = 2.33√P

0.01

0.12 m

0.12 m

0.23 m

0.1

0.37 m

0.37 m

0.74 m

1

1.17 m

1.17 m

2.33 m

10

3.70 m

3.70 m

7.37 m

100

11.7 m

11.7 m

23.3 m

For transmitters rated at a maximum output power not listed above, the recommended 
separation distance (D) in meters (m) can be estimated using the equation applicable to the 
frequency of the transmitter, where P is the maximum output power rating of the transmitter 
in watts (W) according to the transmitter manufacturer.

NOTE 1:

 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies

NOTE 2: 

These guidelines may not apply in all situations. Electromagnetic propagation is 

affected by absorption and reflection from structures, objects and people.

Appendix 2. Warranty and Replacement Program

The warranty period for the aView is one year from delivery to the customer. Ambu agrees  
to replace an aView free of charge if proof can be provided of faulty materials or faulty 
workmanship. In doing so Ambu cannot accept the cost of transportation or risk of shipment. 
Repair of an aView due to faulty materials or faulty workmanship may be offered in situations 
where this option is available. Ambu holds the right to make the choice between repair or 
replacement in every situation. 
Ambu holds the right to ask the technical department at the customer site, or a similar 
department, to carry out repair of an aView under proper guidance by Ambu.
A defective aView must be handled exclusively by persons authorised by Ambu A/S.  
To prevent infection, it is strictly forbidden to ship contaminated medical devices.  
The medical device (aView or Ambu visualization device) must be decontaminated on site 
before shipment to Ambu. The cleaning and disinfection procedures explained in section 5 
shall be followed. Ambu reserve the right to return contaminated medical devices to the sender.

Guidance and manufacturer’s declaration – electromagnetic immunity

The system is intended for use in the electromagnetic environment specified below.  

The customer or the user of the system should assure that it is used in such an environment.

Immunity test

IEC 60601 
test level

Compliance 
level

Electromagnetic  
Environment Guidance

Conducted 
Radio  
Frequency 
IEC 61000-4-6

3 V RMS
0,15 MHz –  
80 MHz
6 V RMS
in ISM bands
80% AM at  
1 kHz

3 V RMS
0,15 MHz –  
80 MHz
6 V RMS
in ISM bands
80% AM at  
1 kHz

Portable and mobile RF 
communications equipment  
should be used no closer to any  
part of the system, including its 
cables, than the recommended 
separation distance calculated  
from the equation applicable to  
the frequency of the transmitter.

Recommended separation distance
d = 1.17√P
d = 1.17√P 80 MHz to 800 MHz
d = 2.33√P 800 MHz to 2.7 GHz

Where P is the maximum output 
power rating of the transmitter  
in watts (W) according to the 
transmitter manufacturer and d  
is the recommended separation 
distance in metres (m).

Field strengths from fixed RF 
transmitters, as determined by  
an electromagnetic site survey,  
a) should be less than the compliance 
level in each frequency range  
b). Interference may occur in the 
vicinity of equipment marked with 
the following symbol.

Radiated 
Radio  
Frequency 
IEC 61000-4-3

3 V/m 80 MHz 
to 2.7 GHz 
80% AM at  
1 kHz

3 V/m 
80-2700 MHz
80% AM at  
1 kHz

NOTE 1: 

At 80 MHz, the higher frequency range applies.

NOTE 2:

 These guidelines may not apply in all situations. Electromagnetic propagation is 

affected by absorption and reflection from structures, objects and people.

a) Field strengths from fixed transmitters, such as base stations for radio (celluar/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV 
broadcast, cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey shall be considered. 
If the measured field strength in the location in which the system is used exceeds the 
applicable RF compliance level above, the system shall be observed to verify normal operation. 
If abnormal performance is observed, additional measures may be necessary, such as 
reorienting or relocating the system. 

b) Over the frequency range 150kHz to 80MHz, field strengths shall be less than 3 V/m.

16

17

Summary of Contents for aView JANUS2-W08-R10

Page 1: ...For use by trained clinicians physicians only For in hospital use For use with Ambu visualization devices Further instruction details are available online www ambu com Instruction for Use Ambu aView...

Page 2: ...5b aScope 2 aScope 3 aScope 4 Broncho 6 Ambu is a registered trademark and aScope and aView are trademarks of Ambu A S Pat Pending Contents Page English Instruction for Use 4 17 Deutsch Bedienungsanl...

Page 3: ...itor 1 1 Intended Use The aViewTM monitor is a non sterile reusable digital monitor intended to display live imaging data from Ambu visualisation devices 1 2 Warnings and Cautions Failure to observe t...

Page 4: ...Reference Number 2 3 aView Parts No Part Function Material 1 Monitor Casing PC ABS Rubber 2 Touch Screen Displays the image from the Ambu visualization device and a touch screen interface PET Glass 3...

Page 5: ...INMETRO Certificate Medical Electrical Equipment ABNT NBR IEC 60601 1 ABNT NBR IEC 60601 1 2 ABNT NBR IEC 60601 2 18 4 Use of aView Numbers in gray circles below refer to illustrations on page 2 4 1 P...

Page 6: ...nsequences by the reprocessor to ensure that the device continues to fulfil its intended purpose Clean and disinfect the aView according to good medical practice using one of the procedures below Clea...

Page 7: ...w file management tab Return to live image by pressing the blue Live image tab or restart aView by pressing the power button for at least 2 seconds When aView is off restart by pressing power button o...

Page 8: ...xternal monitor is displaying the correct input The external monitor shall be capable of receiving NTSC or PAL Toggle between NTSC and PAL in the video out menu on aView for best result The external m...

Page 9: ...n site before shipment to Ambu The cleaning and disinfection procedures explained in section 5 shall be followed Ambu reserve the right to return contaminated medical devices to the sender Guidance an...

Page 10: ...g der Ausstattung oder eine Abschirmung des genutzten Raums erforderlich Anweisungen zur Platzierung des aView finden Sie in den Tabellen in Anhang 1 12 Achten Sie darauf ob es sich beim angezeigten B...

Page 11: ...r die Composite Videoaus gangs Schnittstelle siehe Pos 6 PE PVC Abk rzungen PC Polycarbonat PE Polyethylen ABS Acrylnitril Butadien Styrol ZDC3 Zinklegierung PVC Polyvinylchlorid POM Polyoxymethylen...

Page 12: ...enn der Akkuladezustand sehr niedrig ist sind die Aufnahmefunktionen nicht verf gbar Falls der aView ausgeschaltet und der Akkuladezustand sehr niedrig ist blinkt die Einschalttas te alle zehn Sekunde...

Page 13: ...ischen Sie den aView anschlie end mit einer sterilen in gefiltertem Trinkwasser RO DI angefeuchteten Gaze ab Stellen Sie sicher dass alle R ckst nde des Reinigungs mittels entfernt wurden 7 Wiederhole...

Page 14: ...zstecker geerdet 4 Europ ische Konfiguration CEE 7 Netzstecker geerdet 5 D nische Konfiguration 2 5a Netzstecker geerdet 6 Konfiguration f r die Schweiz Typ J Netzstecker geerdet Weitere Informationen...

Page 15: ...on in der N he befindlichen elektro nischen Ger ten ist unwahrscheinlich HF Emissionen CISPR 11 Klasse B Das System ist f r die Verwendung in allen Einrichtungen geeignet einschlie lich solchen in h u...

Page 16: ...betragen Empfohlene Trennungsabst nde zwischen tragbaren und mobilen HF Kommunikationsger ten und dem System Das System ist f r den Betrieb in einer elektromagnetischen Umgebung bestimmt in der die H...

Page 17: ...Ambu A S Baltorpbakken 13 DK 2750 Ballerup Denmark T 45 72 25 20 00 F 45 72 25 20 50 www ambu com 492 4050 00 V01 2017 09 TCC 10368...

Reviews: