Plugs

Between the power supply 
and aView

Ø5.5mm DC jack connector

6 interchangeable types

1.  Model NEMA 5 AC grounded power plug
2.   Australian configuration: AS3112, AC grounded  

power plug

3.  UK configuration: BS1363, AC grounded power plug
4.   European configuration: CEE 7, AC grounded  

power plug

5.  Danish configuration: 2-5a, AC grounded power plug
6.   Swiss configuration: Type J, AC grounded power plug

Contact Ambu for further information.

8. Trouble Shooting

If problems occur with the aView monitor, please use this trouble-shooting guide to identify 
the cause and correct the error. 

Problem

Possible cause

Recommended action

No live image

on the screen

but User

Interface is

present on the

display or the

image shown 

is frozen.

An Ambu visualization 

device is not connected 

to aView.

Connect an Ambu visualization device to 

the blue or green port on aView.

aView and the Ambu 

visualization device 

have communication 

problems.

Restart aView by pressing the power  

button for at least 2 seconds. When aView 

is off restart by pressing the power button 

once more.

The Ambu visualization 

device is damaged.

Replace the Ambu visualization device with 

a new one.

A recorded image is 

shown in the yellow 

file management tab.

Return to live image by pressing the blue 

Live image tab or restart aView by pressing 

the power button for at least 2 seconds.  

When aView is off restart by pressing 

power button once more.

Low picture 

quality.

Light reflecting on the 

aView screen.

Move aView to a position where no direct 

light influences the screen.

Dirty / damp screen.

Wipe the screen with a clean cloth.

Brightness and contrast 

settings not optimal.

Adjust the contrast and brightness using 

the designated menu on aView.

Battery type

10.8V 4300mAh

Battery Operation

Typical battery runtime of a new, fully charged battery 
(aView turned on and scope connected) is min. 3 hours

Protection against electrical shock

Internally Powered 

Fixture

Mounting interface 

75 mm (2.96”)

Fits poles with thicknesses 

10 mm ~ 45 mm (0.4 ~ 1.8”)

Operating environment

Temperature

10 ~ 40 °C (50 ~ 104 °F)

Relative humidity

30 ~ 85%

IP Protection Classification Sys-
tem

The aView is classified IP30.

Atmospheric pressure

80-109 kPa

Altitude

≤ 2000 m

Dimensions

Width

241 mm (9.49")

Height

175 mm (6.89")

Thickness

33.5 mm (1.32")

Weight

1500 g (331lbs)

7.3. aView Power Supply

Dimensions

Weight

360g (0.79 lbs)

Electrical power

Power requirement

100 - 240V AC; 50-60Hz; 0.6A

Power out

18V DC; 1.67A

Protection against  
electrical shock 

Class I

Operating environment

Temperature

10 ~ 40 °C (50 ~ 104 °F)

Storage

Temperature

10 ~ 40 °C (50 ~ 104 °F)

Relative humidity

10 ~ 90%

12

13

Summary of Contents for aView JANUS2-W08-R10

Page 1: ...For use by trained clinicians physicians only For in hospital use For use with Ambu visualization devices Further instruction details are available online www ambu com Instruction for Use Ambu aView...

Page 2: ...5b aScope 2 aScope 3 aScope 4 Broncho 6 Ambu is a registered trademark and aScope and aView are trademarks of Ambu A S Pat Pending Contents Page English Instruction for Use 4 17 Deutsch Bedienungsanl...

Page 3: ...itor 1 1 Intended Use The aViewTM monitor is a non sterile reusable digital monitor intended to display live imaging data from Ambu visualisation devices 1 2 Warnings and Cautions Failure to observe t...

Page 4: ...Reference Number 2 3 aView Parts No Part Function Material 1 Monitor Casing PC ABS Rubber 2 Touch Screen Displays the image from the Ambu visualization device and a touch screen interface PET Glass 3...

Page 5: ...INMETRO Certificate Medical Electrical Equipment ABNT NBR IEC 60601 1 ABNT NBR IEC 60601 1 2 ABNT NBR IEC 60601 2 18 4 Use of aView Numbers in gray circles below refer to illustrations on page 2 4 1 P...

Page 6: ...nsequences by the reprocessor to ensure that the device continues to fulfil its intended purpose Clean and disinfect the aView according to good medical practice using one of the procedures below Clea...

Page 7: ...w file management tab Return to live image by pressing the blue Live image tab or restart aView by pressing the power button for at least 2 seconds When aView is off restart by pressing power button o...

Page 8: ...xternal monitor is displaying the correct input The external monitor shall be capable of receiving NTSC or PAL Toggle between NTSC and PAL in the video out menu on aView for best result The external m...

Page 9: ...n site before shipment to Ambu The cleaning and disinfection procedures explained in section 5 shall be followed Ambu reserve the right to return contaminated medical devices to the sender Guidance an...

Page 10: ...g der Ausstattung oder eine Abschirmung des genutzten Raums erforderlich Anweisungen zur Platzierung des aView finden Sie in den Tabellen in Anhang 1 12 Achten Sie darauf ob es sich beim angezeigten B...

Page 11: ...r die Composite Videoaus gangs Schnittstelle siehe Pos 6 PE PVC Abk rzungen PC Polycarbonat PE Polyethylen ABS Acrylnitril Butadien Styrol ZDC3 Zinklegierung PVC Polyvinylchlorid POM Polyoxymethylen...

Page 12: ...enn der Akkuladezustand sehr niedrig ist sind die Aufnahmefunktionen nicht verf gbar Falls der aView ausgeschaltet und der Akkuladezustand sehr niedrig ist blinkt die Einschalttas te alle zehn Sekunde...

Page 13: ...ischen Sie den aView anschlie end mit einer sterilen in gefiltertem Trinkwasser RO DI angefeuchteten Gaze ab Stellen Sie sicher dass alle R ckst nde des Reinigungs mittels entfernt wurden 7 Wiederhole...

Page 14: ...zstecker geerdet 4 Europ ische Konfiguration CEE 7 Netzstecker geerdet 5 D nische Konfiguration 2 5a Netzstecker geerdet 6 Konfiguration f r die Schweiz Typ J Netzstecker geerdet Weitere Informationen...

Page 15: ...on in der N he befindlichen elektro nischen Ger ten ist unwahrscheinlich HF Emissionen CISPR 11 Klasse B Das System ist f r die Verwendung in allen Einrichtungen geeignet einschlie lich solchen in h u...

Page 16: ...betragen Empfohlene Trennungsabst nde zwischen tragbaren und mobilen HF Kommunikationsger ten und dem System Das System ist f r den Betrieb in einer elektromagnetischen Umgebung bestimmt in der die H...

Page 17: ...Ambu A S Baltorpbakken 13 DK 2750 Ballerup Denmark T 45 72 25 20 00 F 45 72 25 20 50 www ambu com 492 4050 00 V01 2017 09 TCC 10368...

Reviews: