Section 6 • Patient Lockout
51
NOTE:
The clinician may program one, two, three, or four digits into the pump for the
lock code. If programming only one, two, or three digits into the pump for the
lock code, when the desired code is programmed (e.g., 12 or 123), press and hold
both the RUN/PAUSE button and the BOLUS button for two seconds until one
beep is heard. This will accept the shortened code and lock the pump program.
After completing lockout mode, the pump will return to pause mode and the LOCK icon
will appear in the display. The current infusion can be resumed after reviewing the current
program settings. (See Section 3.3.)
6.3
To Unlock the Pump
To unlock the pump program, the clinician accesses the lockout mode, as described above. The
code previously entered is re-entered into the pump by entering one number into each space of
the display, starting from the right.
Step #1
The first digit space is flashing.
Step #2
Press and release the RUN/PAUSE button to increase the number or press and release the
BOLUS button to decrease the number.
Step #3
When the desired number appears in the display, press and hold either the RUN/PAUSE
button or the BOLUS button for two seconds until one beep is heard.
Step #4
Repeat Steps 1 through 3 to enter the second through fourth digits.
NOTE:
If entering only one, two or three digits for the lock code, when the code is
entered (e.g., 12 or 123), press and hold both the RUN/PAUSE button and the
BOLUS button for two seconds until one beep is heard. This will accept the
shortened code and unlock the pump program. After unlocking, the pump will
return to pause mode. The clinician can now reprogram the pump, clear the
history, or verify and resume the current program settings.
NOTE:
If a code is forgotten, contact Summit Medical Products, Inc. to obtain technical
support. (See Section 13.)
NOTE:
The pump program must be unlocked for programming, reprogramming and
clearing the infusion history.