Section 3 • Programming Instructions
38
NOTE:
When programming P+P mode, if the tens and/or ones digits are already set to
the desired setting, press the RUN/PAUSE and the BOLUS buttons at the same
time until a beep is heard to accept the setting and move on to the next step. This
option is not available when a parameter is programmed to its maximum value.
3.2.3.2
P+P Mode Programming Steps
NOTE:
If at any point during programming the user needs to move backward in the
display, refer to instructions found at 3.2.3.3.
ACTION
AUDIBLE
INDICATOR
VISUAL INDICATOR
1
Enter program mode:
From pause mode, press and hold
both the RUN/PAUSE button and
the BOLUS button until one beep
is heard.
One beep
“PRO” is displayed for two
seconds to indicate that the
pump is in program mode
2
Select P+P:
Press and release the RUN/PAUSE
button to scroll up or press and
release the BOLUS button to scroll
down through the program modes
until “P+P” appears in the display.
Press and hold the RUN/PAUSE
button until one beep is heard to
accept the setting
One beep
“P+P” flashes in the display.
Mode is established.
3
STEP 3 APPLIES ONLY TO THE PIB-IL
v1.5
PUMP: If programming the PIB*PCA
v1.5
, skip Step 3
and proceed to Step 4.
Set interval limit (“IL”):
(A) Press the RUN/PAUSE button
to increase the number or press
the BOLUS button to decrease the
number.
(B) When the desired number
appears in the display, press and
hold either the RUN/PAUSE button
or the BOLUS button until one beep
is heard to accept the setting.
One beep
Letters “IL” in the display;
numbers and “ml” flash.
Interval limit is established.
NOTE: Dose volume can be programmed only to the maximum of the IL setting.
NOTE: Bolus volume can be programmed only to the IL setting minus the dose volume setting.
ml