Section 3 • Programming Instructions
30
ACTION
AUDIBLE
INDICATOR
VISUAL INDICATOR
4
Set bolus volume:
(A) Press and the RUN/PAUSE
button to increase the number or
press and release the BOLUS button
to decrease the number.
Letter “b,” numbers left of
decimal point, “ml” and
bolus icon flash.
(B) When the desired number
appears in the display, press and
hold the RUN/PAUSE button until
one beep is heard to accept the
setting.
One beep
NOTE: Bolus volume is
displayed in ml.
(C) Repeat the steps 4(A) and 4(B)
to set the digit to the right of the
decimal point.
Letter “b” is displayed and
the number to the right of
decimal point, “ml,” and
bolus icon flash. Bolus
volume is established.
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS
button is pressed, the
display moves backward
one screen to enable review
or correction.
NOTE
: The limits for bolus volume are 0.0 to 50.0 ml.
5
Set lockout time:
(A) Press and release the RUN/
PAUSE button to increase the time
or press and release the BOLUS
button to decrease the time.
“LOt” and symbols are
displayed for two seconds
to indicate that the pump is
in lockout mode, “hr:min” is
displayed and the bolus,
lock and clock icons, as well
as the numbers in the hr
section will flash.
(B) When the desired lockout time
appears in the display, press and
hold the RUN/PAUSE button until
one beep is heard to accept the
setting.
One beep
NOTE
: Lockout time is
displayed in hours and
minutes.
(C) Repeat the steps 5(A) and 5(B) to
set the digits in the minutes section.
“hr:min” is displayed and
bolus, lock and clock icons,
as well as numbers in the
min section, flash.
Lockout time is established.
ml
ml