Section 7 • Alarms and Signals
52
SECTION 7 - ALARMS AND SIGNALS
PRIORITY
STATUS
ICON
VISUAL
INDICATOR
AUDIBLE
INDICATOR
COMMENTS
Low
Pump is
infusing
normally
Green run light
(inside the BOLUS
button) blinks,
“ml” icon and
volume infused is
displayed
None
Periodic movement
of the cassette gears is
normal.
Low
Bolus
infusion
Green run light
(inside the BOLUS
button) double
blinks; bolus icon
is in the display
One beep
One beep will sound
every time the BOLUS
button is pressed during
run mode.
Low
Infusion
paused
Pause icon flashes
in the display;
green run light
(inside the BOLUS
button) is off
Two beeps
every four
minutes
Low
Low battery
Battery icon flashes
in the display
Five short
beeps every
four minutes
The battery icon will
remain
flashing in the
display. Replace the
batteries as soon as
possible.
Low
Dead
battery
Battery icon and
alarm icon are in
the display
Constant tone
Press the RUN/PAUSE
button to silence the
alarm. The alarm and
battery icons will remain
displayed. Replace the
batteries immediately.
Low
Cassette not
attached to
pump
Alarm icon and
“CASS” are in the
display
Constant tone
Press the RUN/PAUSE
button to silence the
alarm. Gently press on
top of the cassette to
ensure proper placement.
Resume infusion.
Low
BOLUS
button
None
“REL” is in the
display
Constant tone Release the BOLUS
button