Section 12 • Electromagnetic Immunity (EMC)
66
Table 12-1. Electromagnetic immunity levels tested and passed by ambIT
®
pump.
Phenomenon
Basic EMC Standard Or Test
Method
Immunity Test Levels
Electrostatic discharge
IEC 61000-4-2
±2, 4, 6, 8 kV contact
±2, 4, 8, 15 kV air
Radiated RF EM fields
IEC 61000-4-3
10 V/m
80 MHz-2.7 GHz
Rated power frequency magnetic
fields
IEC 61000-4-8
30 A/m
50 Hz and 60 Hz
Table 12-2. Maximum measured radiated emission levels from the ambIT
®
pump during
operation.
Frequency (MHz)
dB(µV/m)
Quasi-peak (limit)
Actual value (pass/fail)
30 to 230
30
23.85 (pass)
230 to 1000
37
29.31 (pass)