Section 3 • Programming Instructions
42
•
If the leftmost value is blinking and the BOLUS button is pushed and held until a beep
is heard, then the value is accepted (i.e., stored in memory) and the display changes the
preceding parameter. For example if the left most value of the bolus lockout time (hours)
is blinking and the BOLUS button is pushed and held held until a beep is heard, the hours
value is accepted and the display changes to the previous parameter (bolus volume) and
the right most digit blinks (tenths digit).
•
If the basal rate left most digit (tens digit: 0,1X, 2X) is blinking and the BOLUS button
is held until a beep is heard, then the value is accepted and the pump shows the mode
selected “PCA”.
•
If the mode (PCA) is blinking and the BOLUS button is held until a beep is heard, the
mode is accepted and the pump displays the first parameter for that mode and the left
most value is blinking.
NOTE:
If the mode is blinking, then the BOLUS button functions the same as the RUN/
PAUSE button.
3.3 Program Review
CAUTION:
After programming the pump, always verify the program by performing a
program review.
NOTE:
After turning on, programming, locking or unlocking the pump, the RUN/
PAUSE button may also be pressed and immediately released to review the
program. After this initial review (“RX”), pressing and releasing the RUN/
PAUSE button will start/resume the infusion.
NOTE:
The Remote BOLUS Switch may be used for Program Review.
To review the pump program:
S
tep
1:
Place the pump in pause mode.
S
tep
2:
Press and release the BOLUS button.
“RX” will appear in the display and the pump will scroll through each programmed parameter.
After the review, the pump will return to
pause
mode. Pressing the RUN/PAUSE button will
start or resume the infusion.