Section 3 • Programming Instructions
40
ACTION
AUDIBLE
INDICATOR
VISUAL INDICATOR
6
Set bolus volume:
(A) Press and release the RUN/
PAUSE button to increase the
number or press and release the
BOLUS button to decrease the
number.
Letter “b,” numbers left of
decimal point, “ml” and bolus
icon flash.
(B) When the desired number
appears in the display, press and
hold the RUN/PAUSE button until
one beep is heard to accept the
setting.
One beep
NOTE:
Bolus volume is
displayed in ml.
(C) Repeat steps 5(A) and 5(B) to set
the digit to the right of the decimal
point (00.0 to 00.9).
Letter “b” is displayed, and
number right of decimal point,
“ml” and bolus icon flash.
Bolus volume is established.
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS button
is pressed, the display moves
backward one screen to enable
review or correction.
NOTE
: The limits for the bolus volume are 0.0 to 50.0 ml.
7
Set lockout time:
(A) Press and release the RUN/
PAUSE button to increase the
lockout time or press and release
the BOLUS button to decrease the
lockout time.
“LOt” is displayed for two
seconds to indicate that the
pump is in lockout mode.
“hr:min” is displayed, and
bolus, lock and clock icons, as
well as numbers in the hour
section, flash.
(B) When the desired lockout time
appears in the display, press and
hold the RUN/PAUSE button until
one beep is heard to accept the
setting.
One beep
NOTE:
Lockout time is
displayed in hours and minutes.
NOTE:
If bolus volume is set
to “0” this step will be skipped.
(C) Repeat steps 6(A) and 6(B) to set
the digits in the minutes section.
“hr:min” is displayed, and
bolus, lock, and clock icons, as
well as numbers in the minutes
section, flash.
Lockout time is established
ml