Section 4 • Operating Instructions
43
SECTION 4 - OPERATING INSTRUCTIONS
NOTE:
All operating instructions are the same for the three modes of the ambIT
®
PIB
v1.5
pumps except “Start Infusion” (Section 4.1). Please read respective sections
within Section 4.1 to understand how each mode starts infusion.
NOTE:
Pump display may not be clearly visible in bright light. Shading the display will
allow viewing of the display by the user.
4.1
Start Infusion
NOTE:
Before starting infusion, the pump must be
programmed. See Section 3 - “Programming
Instructions” for details.
To begin an infusion, with the pump in pause mode, press and release
the RUN/PAUSE button. The RUN/PAUSE button is located directly
below the pump display.
The first time an infusion is started after programming, or after
powering on, the pump will review the program settings (“RX”) when
either the RUN/PAUSE button or the BOLUS button is pressed. After
this review, the pump will return to pause mode.
Press and release the RUN/PAUSE button again to start the infusion.
The green run light (inside the BOLUS button) will start to blink, the
“ml” icon and the volume infused will appear in the pump display.
Please refer to Sections 4.1.1, 4.1.2, and 4.1.3 for instructions on
receiving infusions for different modes.
4.1.1
PCA Mode Start Infusion
The bolus can be delivered immediately by pushing the BOLUS button
after placing the pump in run mode after changing the setting or
clearing the history. The lockout time is from the end of a bolus to the
start of the next bolus.
4.1.2
PIb Mode Start Infusion
The first PIb dose is delivered after the dose interval with the following
exception:
Pump
display
RUN/
PAUSE
button
BOLUS
button