Section 6 • Patient Lockout
50
SECTION 6 - PATIENT LOCKOUT
The pump has been designed with patient lockout features:
WARNING:
The steps required for entering the pump’s program mode should not be
revealed to the patient. This prevents unauthorized tampering with the infusion
parameters, and prevents the infusion history from being cleared.
6.1
Accessing Lockout Mode
The pump’s lockout mode is accessed by following these steps:
Step #1
Place the pump in pause mode.
Step #2
Enter program mode by pressing and holding both the RUN/PAUSE button and the
BOLUS button for two seconds until one beep is heard and “PRO” appears in the pump
display.
Step #3
Release both buttons and immediately press and release both the RUN/PAUSE button
and the BOLUS button again. Four dash symbols (
- - - -
) will appear in the display.
6.2
To Lock the Pump
After accessing lockout mode (see above), enter one lock code number in each space of the
display, starting from the right. The steps are outlined below:
Step #1
The first digit space is flashing.
Step #2
Press the RUN/PAUSE button to increase the number or press the BOLUS button to
decrease the number.
Step #3
When the desired number appears in the display, press and hold either the RUN/PAUSE
button or the BOLUS button for two seconds until one beep is heard.
Step #4
Repeat Steps 1 through 3 to program the second through fourth digits.