Data logger
15. DATA LOGGER
Measuring instrument ALMEMO
®
1036 can be
made into a data logger - by fitting memory
connector ZA1904SD with a micro-SD memory
card (available as an accessory). In the main
menu an additional menu item
Data logger
LOG
should then appear automatically; with this
you can call up two further sub-menus offering
a comprehensive range of data logger func-
tions. These can be used to acquire measured values either manually at
specifiable points in time or cyclically over a specifiable period and record
these on the memory card.
15.1 Memory connector with memory card
Measured data is written to this memory card via the memory connector; this
data is in table mode and standard FAT16 format. The memory card can be
formatted and its contents can be read out and / or deleted on any normal PC
using any standard card reader. This data can also be imported into MS-Excel
or into Win-Control.
The memory connector fitted with a memory
card can be plugged in at socket A2; it will be
recognized automatically. The first data logger
menu should then be accessible - with date,
time-of-day, cycle, and file name. It should also
be possible to view the memory properties - its
total memory capacity, memory still free, and memory time available.
Card memory - total capacity
External memory 128.00 MB
Card memory - still free
Memory capacity free 21.75 MB
File name (maximum 8 characters)
File name
.001
The
device status can be checked
by means of the symbols appearing in the
top status bar of the menu. (see 9.2)
Before starting any measuring operation you can, in the
File name
function, en-
ter an 8-character file name. In the absence of a user-defined file name, the
default ´ALMEMO.001´ or the name most recently used will be suggested auto-
matically. So long as the connector configuration is not altered, any number of
measuring operations can be saved - either manually or cyclically - all in the
same file.
If, however, the
connector configuration
has been
changed
since the last
measuring operation, a new file will be created; and, if no new file name has
been programmed, the index in the file name extension will be incremented au-
tomatically by 1, e.g. ´.002´. Similarly, if the file name now entered already ex-
ists, a new file will be created with the same file name prefix but with a new in -
dex.
ALMEMO
®
1036
29
© REC COM l© ©l * ´´´´´´µµµµ¶
Time: 12:34:56 Date: 01.01.06
Cycle-Timer:
00:00:30 S
Memory Ext.:
250.0 MB
Memory free:
250.0 MB
Memory duration: 24T 13h
Filename:
ALMEMO.001
STOP Mªª © F MANU
ALMEMO 1036
M*
Sensor display ©
M Measuring points list
M Max-Min, individual mem.
M Datenlogger LOG
P Sensor programming
P Device Configuration
INFO Mªª F © *ON