8. Connecting sensors / transducers
2. Ensure that the humidity sensor is always sufficiently moistened. If you are
unsure whether there is sufficient moisture carry out a visual inspection of
both the cotton sheath and the wick itself. To moisten the wick use distilled
water only. Ordinary water may leave calcified deposits on the wick.
3. If the wick no longer absorbs water properly (through dirt or dryness), re-
place its cotton sheath.
4. Air velocity at the air intake must be minimum 2 m/s. Ensure therefore that
the air intake is always free and unobstructed.
5. Avoid inadvertently warming the probe head from external heat sources or
your own body heat.
8.3 Measuring sockets, measuring channels
Measuring instrument ALMEMO
®
1036 has 2 measuring sockets M0 and M1
(1) and to each of these up to 4 measuring channels can be allocated.
Device-internal channels
This device also provides 4 device-internal additional channels. These can be
used, as and when necessary, to program the battery voltage, the device-inter-
nal atmospheric pressure, or a function channel Differential . see 14.3, 11.3.
Channel assignment
on the 1036-2:
The 3 possible sensors are programmed on leaving
our factory with the following ranges (s. 14.3):
Pt100-Psychrometer
FPA836-3P3
Psychrometer
FNAD46-3
Humidty sensor
FHAD36-Rx
4.
D r
D AP
D AP
3.
P2td
D Uw
D td
2.
P2Uw
D tw
D Uw
1. P314 HT P314 TT AP
D t
D t
M0
M1
M2 M0/M1
M0/M1
16 ALMEMO
®
1036
1. chan.
2. chan.
3. chan.
4. chan.
sensor channels
1036-2
00
10
20
30
01
11
21
31
02
12
22
32
M0 M1 M2
internal
channels