Functions of the Pt100 psychrometer FPA8363P3
tion can be stored in the sensor and interpolated on a linear basis for subse-
quent measuring operations.
14.4.3 Smoothing by means of a sliding average
When measuring temperature at a resolution of 1/1 000 °C - especially if using
sensors in air - the measured value display may be unstable. In this case the
measured value can be smoothed in the form of a sliding average over a speci-
fied time frame. The
level of smoothing
can be set in the
Smoothing
function
by specifying the number of measured values to be averaged (range 0 to 99);
this is for both Pt100 channels.
Measured value smoothing over e.g. 15 values
Smoothing : 15
Time constant
t
100
(s) = smoothing / (conversion rate * 2) = 15 / (1.25 * 2) = 6 seconds
14.5 Functions of D6 sensors
Normally the functions and ranges of digital D6 sensors can only be config-
ured on a PC connected via an interface cable. However, with this special de-
vice - an exception - the measuring ranges and their abbreviations and the at-
mospheric pressure compensation can also be programmed in the sensor it-
self; it can subsequently be used in the modified form in other devices too.
However, multi-point calibration and time constant can only be entered via a
PC. (see D6 instructions).
14.5.1 Measuring ranges of the NTC psychrometer FNAD46-3
Measuring function
Sensor
Meas. Range Unit
Display
Command
*Ntc-Dry Temperature TT
FNAD46-3
-50.00...+125.00 °C
D TT
B-O1
*Ntc-Humid Temperature HT FNAD46-3
-50.00...+125.00 °C
D HT
B-O9
*Rel. Atm. Humidity 0.1%H
FNAD46-3
0.0... 100.0 %H
D rH
B-O2
*Atm. Pressure AP in sensor
FNAD46-3
300.0... 1100.0 mbar
D AP
B-O8
Dew-point Temperature 0.1°C FNAD46-3
-64.80... 100.00 °C
D DT
B-O3
Mixture MH, r
FNAD46-3
0.0... 6500.0 g/kg
D MH
B-O4
Abs. Humidity AH, dv
FNAD46-3
0.0... 596.3 g/m
3
D AH
B-O5
Vapor pressure VP, e
FNAD46-3
300... 1100.0 mbar
D VP
B-O6
Enthalpy En, h
FNAD46-3
0.0... 6500.0 kJ/kg
D En
B-O7
14.5.2 Meas. ranges of the capacitive humidity sensor FHAD36-RS
Measuring function
Sensor
Meas. Range Unit
Display Command
*Air temperature
FHAD36-RS
-50.00...+125.00 °C
D °C
B-O1
* Rel. Atm. Humidity 0.1%H
FHAD36-RS
0.0... 100.0 %H
D rH
B-O2
*Dew-point Temperature 0.1°C FHAD36-RS
-64.80... 100.00 °C
D DT
B-O3
ALMEMO
®
1036
27
M
=
∑
i
m
i
/
N
m1
m15
time windows