Calibrator Output Operations
2
U1401B User’s and Service Guide
39
used to indicate the start or end position. The last three
digits of the primary display are used to indicate the
number of steps (the number of steps from start to end).
1
Press
MODE
to cycle through position (start or end),
number of steps, and amplitude adjustment. The digit to
be adjusted will flash on the display.
•
For adjusting the amplitude, to adjust the value of the
selected digit. The amplitude can be set to any value
within the selected output range (±1.5 V or ±15 V for
constant voltage output, ±25 mA for constant current
output).
•
For adjusting the number of steps, press
and
to
select the digit to be adjusted, then press
and
to adjust the value of the selected digit. The number of
steps can be set within the range of 0 to 999 steps.
•
Press
for more than one second to directly reset
the time interval and amplitude of the present step to
zero.
2
Press
OUTPUT
to save the settings.
Summary of Contents for U1401B
Page 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Page 16: ...XVI U1401B User s and Service Guide Contents ...
Page 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Page 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Page 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Page 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Page 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Page 146: ...126 U1401B User s and Service Guide 6 Maintenance ...