Performance Tests and Calibration
7
U1401B User’s and Service Guide
145
After finishing the
CAL-0
and
CAL-1
calibration procedures,
the instrument will automatically enter the 1.5 V output
calibration mode.
Output voltage calibration
Follow the steps below to perform calibration for the output
voltage ranges and values listed in
Table 7- 6
:
1
As you enter each calibration step, the primary and
secondary displays show the
output voltage value
and
“- rdy- ” respectively.
2
Press
OUTPUT
. The primary and secondary displays show
the
output voltage value
and “00000” respectively, which
means the present output level is as shown on the
primary display.
3
Press
or
to adjust the output voltage until the
multimeter reading is the same as the value shown on the
primary display.
4
Press
MODE
to enter the next calibration step.
At the end of the last calibration step, the primary display
will show “PASS” after the
MODE
button is pressed.
Table 7-6
Output voltage calibration steps
Voltage range
Calibration
step
Output voltage value
1.5 V
1
+0.0000 V
2
+1.1000 V
3
–1.1000 V
15 V
4
+00.000 V
5
+11.000 V
6
–11.000 V
Summary of Contents for U1401B
Page 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Page 16: ...XVI U1401B User s and Service Guide Contents ...
Page 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Page 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Page 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Page 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Page 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Page 146: ...126 U1401B User s and Service Guide 6 Maintenance ...