136
U1401B User’s and Service Guide
7
Performance Tests and Calibration
8
Turn the rotary switch to
.
Connect a diode to the
U1401B input terminals in
forward bias position.
2 V
1.9 V
±1.45 mV
9
Turn the rotary switch to
mV
.
Press and hold
for
more than 1 second.
Connect K-type
thermocouple to the
U1401B input terminals.
–40 °C to
1372 °C
0 °C
±3 °C
–40 °F to
2502 °F
32 °F
±6.096 °F
Table 7-3
Input performance verification tests (continued)
Step
Test Function
Connection to 5520A
Range
5520A output
Error from
nominal 1
year