U1401B User’s and Service Guide
2
Calibrator Output Operations
Figure 2-4
Selecting autoramp output mode
Defining the autoramp parameters in the memory
Press and hold
MODE
for more than one second to enter the
autoramp adjustment mode. The ramp function is a dual
slope output. You may adjust the number of steps between
the start and end positions or the end and start positions,
and the amplitudes of the start and end positions.
When the U1401B is in the autoramp adjustment mode, the
secondary display shows the amplitude of the start or end
position. The first digit on the left of the primary display is
Figure 2-5
Ramp output
Press
or
Press
or
Size of each step
= (end amplitude
– start amplitude)
/resolution
Size of each step
= (start amplitude
– end amplitude)
/resolution
Start
position
End position