Performance Tests and Calibration
7
U1401B User’s and Service Guide
143
Temperature calibration
1
In calibration mode, turn the rotary switch to the mV
position.
2
Press
for more than one second to enter
temperature calibration.
3
Plug a K- type thermocouple into the input terminal. Feed
in the reference input signal that represents 0
o
C and
wait for 10 minutes.
4
Press
to finish the temperature calibration.
d.
Turn the rotary switch to
.
Diode
Short
Dual banana plug
with copper wire
shorted between 2
terminals
Short
1.9 V
1.9 V
1.9000 V
e.
Turn the rotary switch to
position.
Press
to select DC.
50 mA
5 mA
5.000 mA
45 mA
45.000 mA
500 mA
50 mA
50.00 mA
450 mA
450.00 mA
f.
Turn the rotary switch to the
position.
Press
to select AC.
50 mA @ 1 kHz
5 mA
5.000 mA
45 mA
45.000 mA
500 mA @ 1 kHz
50 mA
50.00 mA
450 mA
450.00 mA
Table 7-5
Calibration tests
Step
Test function
Cal range
Input
U1401B
Summary of Contents for U1401B
Page 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Page 16: ...XVI U1401B User s and Service Guide Contents ...
Page 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Page 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Page 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Page 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Page 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Page 146: ...126 U1401B User s and Service Guide 6 Maintenance ...