Test Procedures
The test procedures are as follows (for relays closing on alarm event):
n
Put an Ohmmeter on connector pins 1 and 2 of the nurse-call connector.
Generate a
UHG
alarm condition and check if the resistance changes from high resistance to
less than 0.3 Ohm.
n
Put an Ohmmeter on connector pins 3 and 4 of the nurse-call connector.
Generate a
\HOORZ
alarm condition and check if the resistance changes from high resistance to
less than 0.3 Ohm.
n
Put an Ohmmeter on connector pins 5 and 6 of the nurse-call connector.
Generate an
,123
alarm condition and check if the resistance changes from high resistance
to less than 0.3 Ohm.