Fixing Recorder Problems
The V24/V26 Series uses the M1116A and M1116B Recorder Modules to make hard-copy
recordings of the patient data which is displayed on the screen. The method for
troubleshooting is the same for either recorder: identify the physical or functional problem,
then find it in the accompanying table and follow the corrective action that is recommended.
7DEOH7URXEOHVKRRWLQJ5HFRUGHU3UREOHPV
6\PSWRP
3RVVLEOH&DXVH
&RUUHFWLYH$FWLRQ
System thinks that door is
open when it is not.
Module out of paper.
Defective door switch.
Insert new paper roll.
Replace door switch. Exchange module.
System thinks that the
recorder is out of paper
when it is not.
Paper-out sensor dirty or
dusty.
Clean paper-out sensor.
Recorder not
communicating with
System.
Poor connection to the front-
end rack.
Only one recorder module
may be used with each
monitor.
System not configured
properly.
Too many modules
connected.
Unplug the module. Plug it back in and try it
again in a few seconds. (Watch for the LED
to flash.)
Remove one of the recorder modules.
Check the configuration of the connected
monitor.
Check and remove the extra modules.
Recorder won’t run.
Recorder interface not
working correctly.
Unplug the module. Plug it back in and try it
again in a few seconds. (Watch for the LED
to flash.)
Poor print quality.
Printhead dirty.
Printhead failure.
Clean the Printhead as described in Chapter
2 of this manual.
Exchange the module.
Paper not feeding
properly.
Paper roll off center.
Dirty roller.
Center paper roll on roller guides.
Clean roller.
Module does not lock into
rack.
Locking plates defective.
Remove and exchange the locking plates.
Summary of Contents for M1205A V24
Page 1: ...3DUW 1R 0 3ULQWHG XQH 7KLUG GLWLRQ JLOHQW 0 9 DQG 9 6HUYLFH XLGH...
Page 4: ......
Page 12: ...12 Contents...
Page 16: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 2YHUYLHZ...
Page 214: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSDLULQJ WKH JLOHQW 0RGHO 9 Repairing the Agilent Model V24...
Page 301: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSODFHPHQW 3DUWV Replacement Parts 17 4 3 7 12 5 6 26...
Page 358: ...QGH...