5-40 Troubleshooting Plug-in Modules
$JLOHQW0$996HULHV6HUYLFH*XLGH
T
r
ou
b
le
s
ho
ot
in
g P
lu
g
-
in
Mo
du
les
6HOI7HVW3URFHGXUH2YHUYLHZ
6WHS
Choose a patient category (Adult/Pedi/Neo) by pressing the following key sequence:
6WHS
Display 30 seconds of the waveforms and numerics of all connected modules by
pressing:
6WHS
Repeat Step 2 as necessary to check the individual test waveforms and numerics for the
category.
6WHS
Use Step 1 to choose a new patient category and repeat Steps 2 and 3.
See the following table for test values.
7DEOH6\VWHP6HOI7HVW9DOXHV
0RGXOH
7HVW1XPHULF
7HVW:DYHIRUP
ECG
100 bpm in ADULT mode
125bpm in NEO/PED mode
Simulated Square Wave
RESP
15 rpm in ADULT mode
30 rpm in PEDI mode
55 rpm in NEO mode
Simulated Wave
Pressure Systolic
120 mmHg ADULT
60 mmHg PEDI/NEO
Simulated squarewave
Pressure Diastolic
0 mmHg ADULT
0 mmHg PEDI/NEO
Simulated squarewave
SpO
2
100%
No Wave
Pleth
Wave on display
Simulated wave
Cardiac Output
Blood Temp
Inj Temp
5 l/min (+0.1 l/min)
37°C
0°C
No wave
tcpO
2
60 mmHg (8.0 kPa)
No wave
tcpCO
2
40 mmHg (5.3 kPa)
No wave
ETCO
2
40 mmHg (6.0 kPa)
Simulated squarewave
IMCO
2
0 mmHg (0.0 kPa)
Simulated squarewave
AWRR
25 rpm
Temperature
40°C (104°F)
No wave
0RGXOH6HWXS
&RQILJ6HWV
&KDQJH3DW&DWHJ
0RQLWRU6HWXS
7HVW6LJQDOV
Summary of Contents for M1205A V24
Page 1: ...3DUW 1R 0 3ULQWHG XQH 7KLUG GLWLRQ JLOHQW 0 9 DQG 9 6HUYLFH XLGH...
Page 4: ......
Page 12: ...12 Contents...
Page 16: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 2YHUYLHZ...
Page 214: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSDLULQJ WKH JLOHQW 0RGHO 9 Repairing the Agilent Model V24...
Page 301: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSODFHPHQW 3DUWV Replacement Parts 17 4 3 7 12 5 6 26...
Page 358: ...QGH...