The following table depicts the slight Power Supply output voltage differences between the
monochrome Model V24, color Model V24C/V26C, and color battery Model V24/V26.
3RZHU6XSSO\,QSXW9ROWDJH
The power supply uses the following input voltages:
• AC Input: 90 to 250 VAC continuous, 48 to 66 Hz
• Maximum Power Consumption: 150 Volt-Amperes
%DWWHU\3RZHU6XSSO\
Capacity: At least 1 hour on full charge at 25°C (77°F) with two batteries. For example, the
discharge time for the module configuration option #A24 (with ECG/Resp, NBP, SpO
2
,
Pressure, Recorder) is 1 hour. The discharge time for the module configuration option #A22
(with ECG/Resp, NBP, SpO
2
, Pressure) is 1 hour 15 minutes.
Charging time: 4 hours to 90% of full capacity if monitor is off. 16 hours to 90% of full
capacity if monitor is on.
If the Agilent Battery Charger (M1278A) is used, the charging time is 2.5 hours to 90%
capacity.
Monochrome Model V24
Color Model V24C/V26C
Color Model V24CT/V26CT
+5 Volts @ 2.0 Amps, max
+5 Volts @ 2.5 Amps, max
+5 Volts @ 2.8 Amps, max
+12 Volts @ 2.0 Amps, max
+12 Volts @ 0.75 Amps, max
+12 Volts @ 0.65 Amps, max
+60 Volts @ 0.35 Amps, max
+60 Volts @ 0.35 Amps, max
+60 Volts @ 0.3 Amps, max
Summary of Contents for M1205A V24
Page 1: ...3DUW 1R 0 3ULQWHG XQH 7KLUG GLWLRQ JLOHQW 0 9 DQG 9 6HUYLFH XLGH...
Page 4: ......
Page 12: ...12 Contents...
Page 16: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 2YHUYLHZ...
Page 214: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSDLULQJ WKH JLOHQW 0RGHO 9 Repairing the Agilent Model V24...
Page 301: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSODFHPHQW 3DUWV Replacement Parts 17 4 3 7 12 5 6 26...
Page 358: ...QGH...