This section lists error codes with specific Device IDs other than 16400 (4010H). The source
of the fault, and meaning of the error codes listed can only be determined by the
identification of the error code’s specific Device ID.
7DEOH
(UURU&RGHVIRU'HYLFH,'VRWKHUWKDQ
'HYLFH
,'
(UURU&RGHV
3RVVLEOHFDXVH)DXOW
&RUUHFWLYH$FWLRQ
8197
(2005H)
18995 (4A33H)
SDN communication problem
(could be caused by temporary
SCC power failure).
Check SDN connection- Restart the Monitor
after reconnecting.
Check/exchange SDN cable.
Check/exchange System Board
17000
(4268H)
46146 (B442H)
Non-fatal error reported by the
recorder module.
* Ensure that there is paper in the module.
* Ensure that the recorder module door is firmly
shut.
32709
(7FC5H)
20120 - 20128
(4E98H - 4EADH)
Data Transfer Module.
Troubleshoot Data Transfer Module.
32744
(7FE8H)
21035 (522BH)
21107 (5273H)
24138 (5E4AH)
Internal time handling problem.
This error does not imply defec-
tive behavior.
Ignore.
32747
(7FEBH)
3733 (0E95H)
11200 (2BC0H)
20004 (4E24H)
20133 (4EA5H)
Internal time handling problem.
This error does not imply defec-
tive behavior.
Ignore.
32747
(7FEBH)
64026 (FA1AH)
Monitor reboots intermittently.
In Service Mode, execute the extended test. If
the test identifies an error, replace the System
Board.
32752
(7FF0H)
22300 (571CH)
Monitor reboots intermittently.
Replace the System Board.
32752
(7FF0H)
35841 (8C01H)
36353 (8E01H)
Monitor reboots intermittently.
Replace the Display Controller Board or Dis-
play Driver Board.
32753
(7FF1H)
20300 (4F4CH)
Rack_IF communication prob-
lem.
Check Rack_IF connection.
Replace System Board.
32753
(7FF1H)
50497 (C541H)
SDN communication problem
(could be caused by temporary
SCC power failure
1. Check SDN connection- Restart the system
after reconnecting.
2. Check/exchange SDN cable.
3. Replace System Board.
32766
(7FFEH)
3212 (0C8CH)
Out of memory.
Replace System Board.
32767
(7FFFH)
90 (005AH)
Internal Software problem.
Ignore.
Summary of Contents for M1205A V24
Page 1: ...3DUW 1R 0 3ULQWHG XQH 7KLUG GLWLRQ JLOHQW 0 9 DQG 9 6HUYLFH XLGH...
Page 4: ......
Page 12: ...12 Contents...
Page 16: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 2YHUYLHZ...
Page 214: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSDLULQJ WKH JLOHQW 0RGHO 9 Repairing the Agilent Model V24...
Page 301: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSODFHPHQW 3DUWV Replacement Parts 17 4 3 7 12 5 6 26...
Page 358: ...QGH...