Agilent InfinityLab ELSD User Manual
95
8
Maintenance
Updating Detector Firmware
6
Put the detector into Service mode by holding down the rear Flash button
whilst switching on the module.
The front panel will display:
SERVICE MODE
Reboot for normal operation
When in Service mode, the Tera Term displays the bootloader version and
firmware menu.
7
Select the firmware you wish to upgrade/downgrade by pressing the
appropriate number (e.g. press
1
to upgrade the main control firmware).
8
Press
U
to begin the update process.