Agilent InfinityLab ELSD User Manual
47
4
Using the Module
Instrument Controls
Error conditions
The ELS Detector is equipped with several sensors and error checking facilities to
ensure safe operation. If an error is detected, the instrument gives an audible
warning and a visible description of the error condition. In event of any error
condition, the unit defaults into the
STANDBY
mode in which the heaters, light
source, and gas are turned off. A complete list of instrument errors and remedial
actions is given in the troubleshooting section of this manual.
Clearing an Error
Once the source of the problem has been corrected, select
RUN
mode to put the
ELSD back into its operational state. If the problem has not been rectified, the
ELSD will repeatedly error when
RUN
mode is selected.
CAUTION
Solvent flooding the detector
It is strongly recommended that the pump stop from the I/O connector of
the ELSD is connected to the HPLC pump to prevent solvent flooding the
detector should an error occur.