7
Error Information
Module Specific Error Messages
Agilent InfinityLab ELSD User Manual
76
Module Specific Error Messages
These errors are detector-specific.
Internal temperature exceeded lower limit
Error ID: 10
Air temperature inside the instrument is
≤
10 °C
Internal temperature exceeded upper limit
Error ID: 11
Air temperature inside the instrument is >40 °C
On-board Vapor sensor failed
Error ID: 12
The vapor sensor located on the main control board has failed.
Probable cause
Suggested actions
1
The environmental temperature is outside
the specified operating limits of the
instrument.
•
Increase ambient temperature where
detector is located.
•
Move the detector to a warmer location.
Probable cause
Suggested actions
1
The environmental temperature is outside
the specified operating limits of the
instrument.
•
Decrease ambient temperature where
detector is located.
•
Move the detector to a cooler location.
Probable cause
Suggested actions
1
Vapor sensor not connected to the
mainboard.
Please contact your Agilent service
representative.
2
Defective vapor sensor.
Please contact your Agilent service
representative.