Agilent InfinityLab ELSD User Manual
67
6
Troubleshooting and Diagnostics
General Problems
General Problems
Baseline noise
Probable cause
Suggested actions
1
Poor nebulization
Increase the temperature of the nebulizer by
10 °C until the baseline noise decreases.
2
Insufficient evaporation
•
Increase the temperature of the evaporator
by 10 °C until the baseline noise decreases.
•
Increase the evaporation gas flow rate.
•
Decrease the nebulization temperature.
3
Nonvolatile additive in mobile phase
Use a volatile mobile phase.
4
Pressure difference created inside nebulizer
chamber
•
Ensure that the end of the liquid waste tube
is not immersed in liquid.
•
Ensure that the exhaust tube at rear of unit is
not blocked, or extraction is too strong.
5
Pump pulsations, especially in MicroBore
applications where low flow rates are used
•
Use a pulse free pump.
•
Increase the backpressure on the pump by
fitting a backpressure column between the
pump and the injection valve.
•
Use a pulse dampener directly after the
pump in the system.