LAN Configuration
PC and OpenLab CDS, OpenLab ChemStation or OpenLab EZChrom Software
Setup
Agilent InfinityLab ELSD User Manual
40
5
Cooled option is selected by default. Therefore, deselect
Cooled model option
if required.
6
Select
OK
to complete the configuration.
NOTE
The detector’s serial number is the default hostname.
NOTE
If the detector is assigned a temporary IP address (e.g DHCP server),
communication with the software might be prevented following a power cycle of
the instrument.
Recommend using a static IP address to ensure communication with the
software is preserved after power cycle.