Agilent InfinityLab ELSD User Manual
93
8
Maintenance
Updating Detector Firmware
Updating Detector Firmware
The ELSD Firmware can be upgraded using Lab Advisor. Where Lab Advisor is
not available, the following process may be used.
The Agilent 1290 Infinity II ELSD (G7102A) also contains additional firmware on
the Peltier unit and the Laser assembly.
Firmware on all three assemblies can be upgraded using the same process.
Firmware upgrade is only possible via the serial port.
To upgrade/downgrade the modules firmware carry out the following steps:
1
Connect the detector to a PC, via the serial port on the rear of the module,
using a RS-232 cable.
2
Open the Tera Term program, and select
File >New Connection
from the
toolbar.
When
For majority of internal repairs
Tools required
Description
Allen keys
TERA TERM software
Current firmware file
Parts required
Description
Serial cable (supplied with instrument)
NOTE
Only Firmware versions v30.42 or later are compatible with the 1290 Infinity II
ELSD.