Error Information
Module Specific Error Messages
Agilent InfinityLab ELSD User Manual
78
Fan Failed
Error ID: 16
Thermal shut-down of the main cooling fan.
Fan Stopped
Error ID: 17
A main cooling fan in the module has stopped.
Nebulizer temperature limit exceeded
Error ID: 18
Nebulizer temperature exceeded threshold after stabilizing.
Probable cause
Suggested actions
1
Fan cable disconnected.
Please contact your Agilent service representative.
2
Defective fan.
Please contact your Agilent service representative.
3
Defective mainboard.
Please contact your Agilent service representative.
Probable cause
Suggested actions
1
Obstruction of fan blades.
Please contact your Agilent service representative.
2
Defective fan.
Please contact your Agilent service representative.
3
Defective mainboard.
Please contact your Agilent service representative.
Probable cause
Suggested actions
1
Defective thermocouple.
Please contact your Agilent service representative.
2
Defective nebulizer heater.
Please contact your Agilent service representative.
3
Defective mainboard.