![Agilent Technologies InfinityLab 1290 Infinity II User Manual Download Page 81](http://html.mh-extra.com/html/agilent-technologies/infinitylab-1290-infinity-ii/infinitylab-1290-infinity-ii_user-manual_2868509081.webp)
Agilent InfinityLab ELSD User Manual
81
7
Error Information
Module Specific Error Messages
Cooled evaporator current out of range or
communication failed
Error ID: 25
Peltier module current outside of normal range or communication to Peltier
module has failed.
Laser temperature out of range
Error ID: 26
Temperature control on Laser assembly outside of normal operating range.
Laser current out of range
Error ID: 27
Laser current outside of normal operating range.
Probable cause
Suggested actions
1
Defective Peltier assembly.
Please contact your Agilent service representative.
2
Peltier assembly not connected to
mainboard.
Please contact your Agilent service representative.
3
Defective mainboard.
Please contact your Agilent service representative.
Probable cause
Suggested actions
1
Defective Laser assembly.
Please contact your Agilent service representative.
Probable cause
Suggested actions
1
Defective Laser assembly.
Please contact your Agilent service representative.