Chapter 3: Testing Performance
Offset Accuracy Test
27
15
Set the channel 1 offset value to 400.0 mV. This can be done using the front panel control
or:
a
Pull down the Setup menu and select Channel 1 or click the Channel 1 setup icon.
b
Click the Offset control arrows to change the offset value or click on the offset value
and enter 400.0 mV in the dialog box.
c
Enter 400.0 mV in the Enter Offset dialog box.
16
Set the Aux Out voltage (V
Aux Out
) to +400.0 mV as follows:
a
Pull down the Utilities menu and select Calibration.
b
Change the Aux Output function to DC (top left corner).
c
Set the Level to 400.0 mV.
d
Click on Close.
17
Press the Clear Display key on the scope, wait for the #Avgs value (top left corner of
screen) to return to 256 and then record the DMM voltage reading as V
DMM+
and the
scope Vavg reading as V
Scope+
in the Offset Gain Test section of the Performance Test
Record.
Channel 1
setup icon
Summary of Contents for Infiniium 90000 A
Page 4: ...4...
Page 13: ...2 To run the self calibration 15 Calibration...
Page 14: ...14 Calibration This chapter provides self calibration procedures for the oscilloscope...
Page 50: ...Chapter 3 Testing Performance Performance Test Record 50...
Page 58: ...Chapter 4 Troubleshooting Power Supply Trouble Isolation 58 Figure 4 4...
Page 71: ...Chapter 4 Troubleshooting Display Trouble Isolation 71 Display Trouble Isolation...
Page 88: ...Chapter 4 Troubleshooting To check probe power outputs 88...
Page 126: ...Chapter 5 Replacing Assemblies To remove and replace the USB or GPIB port 126...
Page 129: ...Chapter 6 Replaceable Parts Exploded Views 129 Exploded Views Front Frame and Front Panel...
Page 130: ...Chapter 6 Replaceable Parts Exploded Views 130 Fan and Acquisition Assembly...
Page 131: ...Chapter 6 Replaceable Parts Exploded Views 131 Power Supply and PC Motherboard...
Page 132: ...Chapter 6 Replaceable Parts Exploded Views 132 Sleeve and Accessory Pouch...
Page 136: ...Chapter 6 Replaceable Parts Replaceable Parts List 136...
Page 138: ...Chapter 7 Theory of Operation 138 Figure 7 1 Oscilloscope Block Diagram...
Page 146: ...146 Figure 7 6 Acquisition board block diagram...