Chapter 3: Testing Performance
Analog Bandwidth - Maximum Frequency Check
41
16
Change the scope set up as follows:
a
Change the channel vertical sensitivity to 10 mV/div.
b
Reset the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz waveform).
17
Change the generator output as follows:
a
Reset the generator frequency to 50 MHz.
b
Change the amplitude to the value suggested for this sensitivity in Table 3-1.
18
Repeat steps 8, 9, and 10 to measure the reference gain at 50 MHz for this sensitivity.
19
Repeat steps 11, 12, 13, and 14 to measure the gain at maximum frequency for this
sensitivity.
20
Repeat steps 15 to 19 to complete measuring gains for remaining sensitivities for
channel 1 in the Analog Bandwidth - Maximum Frequency Check section of the
Performance Test Record.
21
Move the splitter to channel 2 and change the scope configuration as follows:
• Ensure Channel 2 is displayed and all other channels are turned off.
• Set the vertical sensitivity of channel 2 to 5 mV/div.
• Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz waveform).
• Right click on the V rms measurement at the bottom of the screen. When the RMS voltage
measurement setup window is displayed, change the source from Channel 1 to Channel 2.
22
Repeat steps 7 to 20 to complete measuring gains for channel 2.
23
Move the splitter to channel 3 and change the scope configuration as follows:
a
Ensure Channel 3 is displayed and all other channels are turned off.
b
Set the vertical sensitivity of channel 3 to 5 mV/div.
c
Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz waveform).
d
Click on the V rms measurement at the bottom of the screen and select Customize. When
the V rms setup window is displayed, change the source from Channel 2 to Channel 3.
24
Repeat steps 7 to 20 to complete measuring gains for channel 3.
25
Move the splitter to channel 4 and change the scope configuration as follows.
a
Ensure Channel 4 is displayed and all other channels are turned off.
b
Set the vertical sensitivity of channel 4 to 5 mV/div.
c
Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz waveform).
d
Click on the V rms measurement at the bottom of the screen. When the V rms setup
window is displayed, change the source from Channel 3 to Channel 4.
26
Repeat steps 7 to 20 to complete measuring gains for channel 4.
Summary of Contents for Infiniium 90000 A
Page 4: ...4...
Page 13: ...2 To run the self calibration 15 Calibration...
Page 14: ...14 Calibration This chapter provides self calibration procedures for the oscilloscope...
Page 50: ...Chapter 3 Testing Performance Performance Test Record 50...
Page 58: ...Chapter 4 Troubleshooting Power Supply Trouble Isolation 58 Figure 4 4...
Page 71: ...Chapter 4 Troubleshooting Display Trouble Isolation 71 Display Trouble Isolation...
Page 88: ...Chapter 4 Troubleshooting To check probe power outputs 88...
Page 126: ...Chapter 5 Replacing Assemblies To remove and replace the USB or GPIB port 126...
Page 129: ...Chapter 6 Replaceable Parts Exploded Views 129 Exploded Views Front Frame and Front Panel...
Page 130: ...Chapter 6 Replaceable Parts Exploded Views 130 Fan and Acquisition Assembly...
Page 131: ...Chapter 6 Replaceable Parts Exploded Views 131 Power Supply and PC Motherboard...
Page 132: ...Chapter 6 Replaceable Parts Exploded Views 132 Sleeve and Accessory Pouch...
Page 136: ...Chapter 6 Replaceable Parts Replaceable Parts List 136...
Page 138: ...Chapter 7 Theory of Operation 138 Figure 7 1 Oscilloscope Block Diagram...
Page 146: ...146 Figure 7 6 Acquisition board block diagram...