Replacing Assemblies
Use the procedures in this chapter when removing and replacing assemblies and parts
in the Agilent Technologies oscilloscopes.
In general, the procedures that follow are placed in the order to be used to remove a
particular assembly. The procedures listed first are for assemblies that must be removed
first.
The graphics in this chapter are representative of the oscilloscope at the time of this
printing. Your unit may look different.
ESD Precautions
When using any of the procedures in this chapter you must use proper ESD precautions.
As a minimum you must place the oscilloscope on a properly grounded ESD mat and
wear a properly grounded ESD wrist strap.
C A U T I O N
AVOID DAMAGE TO THE OSCILLOSCOPE!
Failure to implement proper antistatic measures may result in damage to the oscilloscope.
Tools Required
The following tools are required for these procedures.
• Torx drivers: T10, T20
• Socket wrench: 9/16 inch
• Torque wrench: 3.5 mm (8 in-lbs)
• Medium size (3/16-in) flat-blade screwdriver
C A U T I O N
REMOVE POWER BEFORE REMOVING OR REPLACING ASSEMBLIES!
Do not remove or replace any circuit board assemblies in this oscilloscope while power is applied.
The assemblies contain components which may be damaged if the assembly is removed or
replaced while power is connected to the oscilloscope.
W A R N I N G
SHOCK HAZARD!
To avoid electrical shock, adhere closely to the following procedures. Also, after disconnecting
the power cable, wait at least three minutes for the capacitors on the power supply to discharge
before servicing this oscilloscope. Hazardous voltages exist on the inverter for the display
monitor.
W A R N I N G
SHOCK HAZARD!
Read the Safety information at the back of this guide before performing the following procedures.
Failure to observe safety precautions may result in electrical shock.
W A R N I N G
INJURY CAN RESULT!
Use caution when the oscilloscope fan blades are exposed as they can cause injury.
Summary of Contents for Infiniium 90000 A
Page 4: ...4...
Page 13: ...2 To run the self calibration 15 Calibration...
Page 14: ...14 Calibration This chapter provides self calibration procedures for the oscilloscope...
Page 50: ...Chapter 3 Testing Performance Performance Test Record 50...
Page 58: ...Chapter 4 Troubleshooting Power Supply Trouble Isolation 58 Figure 4 4...
Page 71: ...Chapter 4 Troubleshooting Display Trouble Isolation 71 Display Trouble Isolation...
Page 88: ...Chapter 4 Troubleshooting To check probe power outputs 88...
Page 126: ...Chapter 5 Replacing Assemblies To remove and replace the USB or GPIB port 126...
Page 129: ...Chapter 6 Replaceable Parts Exploded Views 129 Exploded Views Front Frame and Front Panel...
Page 130: ...Chapter 6 Replaceable Parts Exploded Views 130 Fan and Acquisition Assembly...
Page 131: ...Chapter 6 Replaceable Parts Exploded Views 131 Power Supply and PC Motherboard...
Page 132: ...Chapter 6 Replaceable Parts Exploded Views 132 Sleeve and Accessory Pouch...
Page 136: ...Chapter 6 Replaceable Parts Replaceable Parts List 136...
Page 138: ...Chapter 7 Theory of Operation 138 Figure 7 1 Oscilloscope Block Diagram...
Page 146: ...146 Figure 7 6 Acquisition board block diagram...