Chapter 2: Calibration
To run the self calibration
16
Figure 2-1
Calibration Dialog
Box
4
Click Start, then follow the instructions on the screen.
The routine will ask you to do the following things in sequence:
a
Decide if you want to perform the Time Scale Calibration. Your choices are:
• Standard Calibration - Time scale calibration will not be performed. Time scale
calibration factors from the previous time scale calibration will be used and the 10 MHz
reference signal will not be required. The remaining calibration procedure will continue.
• Standard Calibration and Time Scale Calibration - Performs the time scale calibration.
This option requires you to connect a 10 MHz reference signal to channel 1 that meets
the following specifications. Failure to use a reference signal that meets this specification
will result in an inaccurate calibration.
• Standard Calibration and Reset Time Scalie Calibration - Factory time scale calibration
factors will be used. The 10 MHz reference signal will not be required. The remaining
calibration procedure will continue.
b
Disconnect everything from all inputs and Aux Out.
c
Connect the calibration cable from Aux Out to channel 1.
You must use the 54855-61620 cable assembly with two 54855-67604 adapters for all
oscilloscopes except for the DSO90254A which can use a good quality BNC cable. Failure
to use the appropriate calibration cable will result in an inaccurate calibration.
d
Connect the calibration cable from Aux Out to each of the channel inputs as requested.
e
Connect the 50
Ω
BNC cable from the Aux Out to the Aux Trig on the front panel of the
oscilloscope.
f
A Passed/Failed indication is displayed for each calibration section. If any section fails,
check the calibration cables and run the oscilloscope Self Test in the Utilities menu.
Page 87 also has a little section on what to do if a self cal fails in terms of replacing
assemblies.
5
After the calibration procedure is completed, click Close.
Click here to start
calibration
Clear this check
box before starting
calibration
Frequency: 10 MHz ±0.4 ppm = 10 MHz ±4 Hz
Amplitude: 0.2 Vpeak-to-peak to 5.0 Vpeak-to-peak
Wave shape: Sine or Square
Summary of Contents for Infiniium 90000 A
Page 4: ...4...
Page 13: ...2 To run the self calibration 15 Calibration...
Page 14: ...14 Calibration This chapter provides self calibration procedures for the oscilloscope...
Page 50: ...Chapter 3 Testing Performance Performance Test Record 50...
Page 58: ...Chapter 4 Troubleshooting Power Supply Trouble Isolation 58 Figure 4 4...
Page 71: ...Chapter 4 Troubleshooting Display Trouble Isolation 71 Display Trouble Isolation...
Page 88: ...Chapter 4 Troubleshooting To check probe power outputs 88...
Page 126: ...Chapter 5 Replacing Assemblies To remove and replace the USB or GPIB port 126...
Page 129: ...Chapter 6 Replaceable Parts Exploded Views 129 Exploded Views Front Frame and Front Panel...
Page 130: ...Chapter 6 Replaceable Parts Exploded Views 130 Fan and Acquisition Assembly...
Page 131: ...Chapter 6 Replaceable Parts Exploded Views 131 Power Supply and PC Motherboard...
Page 132: ...Chapter 6 Replaceable Parts Exploded Views 132 Sleeve and Accessory Pouch...
Page 136: ...Chapter 6 Replaceable Parts Replaceable Parts List 136...
Page 138: ...Chapter 7 Theory of Operation 138 Figure 7 1 Oscilloscope Block Diagram...
Page 146: ...146 Figure 7 6 Acquisition board block diagram...