Appendix D. Performance Test
Test V. PDL/PDG uncertainty (#020 only)
a
Mount the protection cap on the head in order to keep the
light out.
b
Press
M
ORE
.
c
Press
Z
ERO
and wait while zeroing.
d
Unmount the protection cap and connect the equipment as
shown in Figure D-6.
4
PDL Measurement:
Do not move the patchcord during measurement!
a
Press
R
ESET
.
b
Note the first stable value after the window full marker (|—|)
is shown at the display in the test record.
5
Move the fiber and the optical head to another position (see
Figure D-7 but don’t move the equipment while measuring.
6
Repeat steps 4 and 5 twenty times.
7
Calculate the PDL/PDG uncertainty as described in the test
record.
Figure D-7
Example Drawings
Optical Head
bend diameter
Optical Head
Optical
Head
Note: Don’t form the bend diameter smaller than 4 inches (10 cm) to prevent termin