135
HP-IB Programming
OLA Specific Commands
related commands
:SENSe:POWer:MEASuring:MODe, :SENSe:POWer:UNIT
example
OUTPUT 724;":SENS1:DATA? POW"
ENTER 724;A$
command
:SENSe:DATA? RL
syntax
:SENSe[1|2]:DATA?<wsp>RL
description
This query returns the Return Loss value. It is only applicable to the
Return Loss application.
parameter
1|2
1
Reads head A
2
Reads head B
Default is 1.
response
The measured value in dB.
errors
The message “Wrong application for this command” is returned, if
the present application is not Return Loss.
The message “No head connected” is returned, if there is no optical
head attached to the specified channel.
The message “No valid result possible” is returned, if the value
could not be measured.
example
OUTPUT 724;":SENS1:DATA? RL"
ENTER 724;A$
command
:SENSe:DATA? SR
syntax
:SENSe[1|2]:DATA? SR
description
This query returns either the Splitting Ratio SR or the Polarization
Dependent Splitting Ratio PDSR, depending on the application
currently active. It is only applicable to the applications
•
Coupler Test, and
•
Polarization Dependent Coupler Test.