157
HP-IB Programming
OLA Specific Commands
command
:SOURce:POWer:STATe?
syntax
:SOURce:POWer:STATe?
description
This query returns the current state of the internal laser source.
response
A
boolean
value:
0
The internal source is off.
1
The internal source is on.
related commands
:SOURce:POWer:STATe
example
OUTPUT 724;":SOUR:POW:STAT?"
ENTER 724;A$
command
:SOURce:POWer:WAVelength
syntax
:SOURce:POWer:WAVelength<wsp><source>
description
This command switches the wavelength of the built-in laser source.
It is only applicable to an instrument equipped with the dual source
option. It can be used any time. It is ignored, if the OLA has only a
single source.
parameter
Source
You enter one of the following keywords:
LOWer
Switches to 1310 nm
UPPer
Switches to 1550 nm
related commands
:SOURce:POWer:WAVelength?
example
OUTPUT 724;"SOUR:POW:WAV UPP"
command
:SOURce:POWer:WAVelength?
syntax
:SOURce:POWer:WAVelength?
description
This query returns the current wavelength of the internal source.
response
The wavelength is returned as a floating point number in
exponential format, e.g. 1.55E-6. The unit is meter. The unit is not
included in the response.
related commands
:SOURce:POWer:WAVelength