62
Taking Standard Loss Measurements
Measuring the Return Loss
Return Loss Measurement Setup
1
Connect the optical output to the input of the DUT.
2
If the DUT has no fiber output, attach a patchcord to its output.
3
Wind the fiber or patchcord close to its open end in seven tight
loops around a pen.
This will virtually eliminate the backreflection from the open
end.
Figure 3-6
Setup for Return Loss Measurements
Head Input A
Head Input B
Optical Input
Optical Output
Head Input A
Head Input B
Optical Input
Optical Output
3 dB
Coupler
Pol. Ctrl.
Internal
Source(s)
Signal Processing and Display
Device
under
Test