209
Appendix D. Performance Test
Test VI. Repeatability PDL/PDG
3
Zero the OLA:
a
Mount the protection cap on the head in order to keep the
light out.
b
Press
M
ORE
.
c
Press
Z
ERO
and wait while zeroing.
d
Unmount the protection cap and connect the equipment as
shown in Figure D-8.
4
PDL Measurement:
a
Press
R
ESET
.
b
Note the first stable value after the window full marker (|—|)
is shown at the display in the test record.
5
Repeat step 4 twenty times.
6
Calculate the Repeatability as described in the test record.