Agilent Technologies Helium Leak Detector
80
DRAFT 12
/1
1/17
Audio Set Point
:
Desired
Audio Alarmed Leak
Rate Value (> <) - This value can be
adjusted from 10
-8
to 10
-4
. The
Audio Set Point
triggers with its'
own distinct tone if different from
Set Point 1
.
NOTE
The Audio Setpoint is very useful in the Sniffing Application.
The programmed Audio Set Point value can be programmed
50% < Set Point 1, which alarms the operator that they are
approaching a leak (leak location).
Plugged Probe Alarm
A distinct high frequency tone different from the
Audio & Set Point
1
. This alarm is useful in alerting the operator that the desired
Power Probe/Sniffing Probe
adjustment (pressure) has changed. A
High Limit
can be selected in alerting that too much flow (high test
port pressure) is present. A
Low Limit
trigger is typically the norm
indicating that the
Power Probe/Sniffing
is blocked or the tubing is
kinked or pinched off.
Files
The operator can store up to three setup configurations:
❑
Sniff 1-3
The information icon provides the necessary directions for a suc-
cessful external calibration.
Calibrate
External calibration/Sensitivity check: Follow the step by step pro-
cess.