Agilent Technologies Helium Leak Detector
16
DRAFT 12
/1
1/17
O-ring Care
When removing, checking or replacing O-rings, keep in mind the following:
NOTE
Agilent recommends replacing all O-rings during routine
maintenance or during any maintenance procedure requiring
that O-rings be removed.
NOTE
Due to the effective cleaning nature of VacuSolv solvent and
its residue-free properties, Agilent’ Component and
Spectrometer Cleaning Kit (Part Number 670029096), used in
accordance with the kit instructions, is recommended for
cleaning spectrometer components. The kit can also be used for
fine cleaning of other parts in the leak detector’s vacuum
system such as valves and fittings. No rinsing steps or
high-temperature drying are required following cleaning with
VacuSolv. Although appropriate precautions are advised,
VacuSolv is compatible with most materials and does not
contain toxic chemicals or CFCs (chlorofluorocarbons). Other
acceptable solvents are isopropyl alcohol (IPA) or Dow
Corning
®
OS-20.
CAUTION
Remove O-rings carefully with your fingers. Do not use metal
tools for this task; this prevents scratching of any sealing
surfaces.
❑
Wipe all O-rings clean with a lint-free cloth before
installation to ensure that no foreign matter is present to
impair the seal.
❑
Do not use grease or any other substance on O-rings that
will come in contact with the spectrometer.
❑
Do not use alcohol, methanol or other solvents on O-rings.
Doing so causes deterioration and reduces their ability to
hold a vacuum.
❑
Agilent does not recommend the use of vacuum grease. If
applicable, apply a small amount of Apiezon
®
L grease and
wipe the O-rings shiny dry.