Agilent Technologies Helium Leak Detector
DRAFT 12/1
1
/17
143
Table A-1 lists inquiries that are used to determine internal operating parameters.
All inquiries, except VER, begin with
?
and end with <CR>.
Table A-1 Internal Operating Parameters
Inquiry
Response
?ALL
Responds with four lines. Each begins with a <cr><lf>. The first line
reports the ion chamber value. The second line reports the emission value.
The third line reports the value of the offset variable. The fourth line reports
the gain value.
?AZ<0
Reports the status of the Auto-zero < 0 function, ON or OFF.
?CALOK
Reports the status of the last calibration.
?CL-XFER
Reports the currently stored contra-flow mode crossover pressure.
?CPLOCKOUT
Reports the status of the control panel lockout feature, ON or OFF.
?CURRDATE
Reports the current system date and time in the following format: 3
05/25/2005 04:19:03.35 1 where 3 is the day of the week (0 = Sunday, 1 =
Monday, etc), 05/25/2005 is mm/dd/yyyy (date), 04:19:03.35 is
hh/mm/ss/ms (time) and 1 is PM (0=AM, 1=PM, 2=24 hour).
?DISPLAY
Reports what the display bar graph is set to,
log
or
lin
.
?EMISSION
Reports the emission value.
?EXPONENT
Reports the current manual-mode exponent with a two-character (minus
followed by one digit) number. The exponent is within the least sensitive
range (as reported by ?RANGE) and the three lower (more negative
exponent) ranges. See also INIT-EXPONENT.
?EXTLEAK
Reports the currently stored value of the external standard leak.
?GAIN
Reports the system gain value by returning a three-character number
consisting of a two-digit gain factor with a decimal point after the first digit.
?INTEXT
Indicates which leak, Internal or External has been selected for use during
calibration.
?IOBOARD
Returns the status: 0 = not installed, 1 = installed
?IO_VER
Returns the software revision of the installed I/O PCB
?IONCHAMBER
Reports the ion chamber value.
?LANGUAGE
Reports the language that has been selected to display on the front panel
(0=English, 1=French, 2=German, 3=Korean, 4=Spanish, 5=Japanese,
6=Chinese or Russian depending on software version purchased).