Agilent Technologies Helium Leak Detector
DRAFT 12/1
1
/17
71
NOTE
For Plugged Probe Alarm - High Pressure Test (Sniffing Mode)
see Section 5.1.3.5 “Sniff Method” on page 79.
❑
Off:
The audio sounds for all leak rates above zero.
Set points may be turned ON or OFF (enabled or disabled).
When the Audio Set Point is OFF, audio sounds for all leak
rates above zero. When the audio set point is ON, audio
sounds only for leak rates above the audio set point.
To configure reject set points and the audio set point:
1. Press the blue check for the Reject Set Point.
2. Press the value button associated with the set point.
The data entry pad appears.
3. Enter the set point and press
OK
.
4. Press the < or > soft key to select whether the reject setpoint activates above or below
the programmed value.
5. Press the blue check for the Audio Set Point.
6. Press the value button associated with the audio set point.
The data entry pad appears.
7. Enter the set point and press
OK
.