98
1200 Series Handheld Control Module Users’s Guide
7
Using the Detectors
Lamp Ignition Routine (VWD only)
Lamp Ignition Routine (VWD only)
When the lamp is turned on, the following routine is implemented:
•
1-minute warm-up, within this time the grating position is re-initialized.
Check of wavelength setting at 656 nm emission (must be within 1 nm) at
reference side only (to have no influence from flow cell condition).
If not correct, a message for calibration is displayed. Perform a wavelength
calibration (available within VWD Tests).
•
Check of intensity
•
at 250 nm without cut-off filter inserted,
•
at 250 nm with cut-off filter inserted.
If the intensity check is not OK, the cut-off filter operation is not OK.