1200 Series Handheld Control Module Users’s Guide
151
Running an Injector Program
12
Entering Injector Program Settings
Entering Injector Program Settings
1
Enter the
Analysis
screen.
2
Select the
Settings
button [
F1
].
3
Select LC System from the menu.
4
Select the
Default
button [
F6
] and select the Yes button to load defaults.
5
Press
Esc
to go back to the
Analysis
menu.
6
Select the
Settings
button [
F1
].
7
Select Autosampler from the menu.
8
Select the Inj.
Program
button [
F3
] to display the
Program
screen.
9
For program line 1, select the
Insert
button [
F7
].
10
Select the DRAW function and enter the draw settings as described in
11
Select the
Enter
button [
F7
] to accept entries.
12
Repeat the procedure for the other functions to complete the program.
13
Press the
Esc
key until you are back in the
Settings
screen.